<4. Display and Operations while Connected to GC8000>

4-32

IM 11B06C01-01E

4.3.29

Stream Sequence

To set a stream sequence, select the Stream Sequence command from the Setup menu. The Set

Stream Sequence dialog box appears.

Figure 4.36

Set Stream Sequence dialog box

In this dialog box, you can assign streams of active GCM stream sequence, and enable or

disable the sequence as follows.
(1) Select a stream sequence number (1 to 8) to set the stream sequence. The corresponding

stream sequence settings are displayed.

(2) Enable or disable the sequence. (Select the box to enable, and deselect the box to disable.)

If you deselect a box, the stream number is indicated as “****” automatically and the stream

number is no longer selectable.

(3) From the stream number combo box, you can select up to 31 stream numbers that are

being assigned to the GCM.

(4) After you have changed the settings, click the Change button. A confirmation message

asking “Are you sure you want to send the data?” is displayed in the message box. When

you respond with “Yes”, the settings are changed.

4.4

Chromatogram Window

The Chromatogram window displays chromatograms of GCM currently connected and

chromatograms stored in PCAS. It also re-integrates chromatograms.
One “online startup” Chromatogram window can be displayed and up to four “off-line startup”

Chromatogram widows can be displayed.
“Online startup”
A startup procedure in which communication with the analyzer or PCAS is possible.

Chromatograms can be acquired from the analyzer or PCAS, and re-integration results can be

transferred.
“Off-line startup”
A startup procedure in which communication with the analyzer or PCAS is not established.

Activate with clicking the icon “Chromatogram” on ASET. Only file chromatograms can be

displayed.

5th Edition : Nov. 16, 2011-00