21 saving and deleting chromatogram, Saving and deleting chromatogram -48, Caution – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 95

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-48

IM 11B06C01-01E

4.4.21

Saving and Deleting Chromatogram

Select the Save Saved Chromatogram command in the File menu to display the Save

Chromatogram dialog box. The dialog box consists of three tabs, “Save”, “Delete”, and

“Baseline”.

CAUTION

“Save” is enabled for user level A or higher, and “Delete” and “Baseline” are enabled for user level

C or higher.

Figure 4.56

Save Saved Chromatogram dialog box (left) Save dialog box in tandem display mode

(right)

l

Saving a saved chromatogram

1) In the Save Saved Chromatogram dialog box, select a chromatogram to be saved as

a saved chromatogram from currently displayed chromatograms in the Chromatogram

window.

2) Click the “Save” button. This causes the “Do you want to save?” confirmation message to

appear.

3) Clicking “OK” saves the chromatogram into GC8000.

5th Edition : Nov. 16, 2011-00