<4. Display and Operations while Connected to GC8000>

4-26

IM 11B06C01-01E

4.3.23

Gate Time

To set a gate time, select the Gate Time tab window from the Set Peak dialog box.

Figure 4.30

Set Peak dialog box (Gate Time)

In this dialog box, you can set the peak reference time, gate ON time, and gate OFF time for the

specified stream number, system number, and peak number.
After you have changed the settings, click the Change button. A confirmation message asking

“Are you sure you want to send the data?” is displayed in the message box. When you respond

with “Yes”, the settings are changed.

Note that you can edit the peak reference time only when the following two conditions are

satisfied. Otherwise, the time is shown as “********” and you cannot edit it.
1) If the Auto Tracking is set to “Track Standard”
2) If the Distillation Analysis is set to “No”

If the Process Setting is set to “Calculation” or “Caloric processing”, all of Peak Reference Time,

Gate ON Time, and Gate OFF time are grayed out and you cannot change them.