Help – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 23

Advertising
background image

<2. Ethernet LCD Emulator Window>

2-3

IM 11B06C01-01E

(4) The analyzer server IP address is illegal.

Countermeasure: Click on the Select Server command in the Connect menu to open the Select Server

dialog box and re-enter the correct IP address.

(5) You have chosen the analyzer ID of an analyzer not connected to the analyzer server.

Countermeasure: Choose the correct analyzer ID.

(6) The selected analyzer is in Local mode.

Countermeasure: Change the mode of the analyzer unit’s LCD panel to Remote and try to make the

communication connection again.

(7) Communication connection with the analyzer cannot be achieved temporarily due to causes such as

noise.

Countermeasure: Execute the Connect command from the Connect menu again.

(5) Click the OK button.

The LCD Emulator window appears.

F0203.ai

HELP !

If the analyzer unit’s LCD panel is being operated while the LCD Emulator is being used, a message stating “Exit

command has been received” appears and communication with the analyzer will be disconnected.
To reactivate the LCD Emulator window in this case, wait for panel operation of the analyzer unit to be finished,

then re-execute the communication connection.

5th Edition : Nov. 16, 2011-00