<4. Display and Operations while Connected to GC8000>

4-35

IM 11B06C01-01E

6) File chromatogram

This chromatogram is automatically saved on ASET or manually saved by the user.

7) Differential chromatogram

This chromatogram is the result of difference determined by comparing two chromatograms.

The difference of shifted chromatograms are determined as they are in the shifted status.

If chromatograms are shifted after their difference was displayed as a differential

chromatogram, the differential chromatogram is not updated automatically. In that case,

display the differential chromatogram again.

4.4.4

Active Chromatogram

The “active chromatogram” is the target for shifting, re-integrating, and other operation intended

for chromatogram. In legend display, the current active chromatogram is highlighted in mazarine.
To select the active chromatogram, manually click a chromatogram displayed in the legend area.
A temperature graph or pressure graph can also be set as active. The active graph can be

shifted. However, re-integration operation is not available while the graph is active.

4.4.5

Legend display

Figure 4.39

Example of legend display

This area displays the correspondence between the pen colors and the names of

chromatograms, temperature graphs, and pressure graphs currently displayed.
By selecting or deselecting the check box of each graph name, corresponding graph can be

displayed in the Enlargement Display window.
If a check box is cleared, the corresponding chromatogram name is highlighted in gray, and the

chromatogram is hidden in the Enlargement Display window.
By clicking the DEL button of a graph, the corresponding graph can be deleted completely.

4.4.6

Auto Gain Mode of Chromatogram

If auto gain mode is selected in the enlargement display area, the voltage value of

chromatogram is enlarged based on the gain setting for each peak.

If auto gain mode is enabled, the Y-axis scale is displayed according to the gain setting. If

auto gain mode is disabled, the Y-axis scale is displayed in mV.

For auto gain mode, the gain setting for each peak must be configured. Therefore, a

chromatogram without gain setting capability does not support auto gain mode.

An unevenness may occur on the portion of chromatogram where the gain level changes.

In auto gain mode, the Y-axis scale is displayed according to the gain in the most left portion

where chromatogram is displayed. Also in an enlargement, the gained chromatogram is

displayed.

For auto gain mode, select the auto gain mode in the scale menu.
The following shows an example of display in auto gain mode.

5th Edition : Nov. 16, 2011-00