<5. Display and Operations while Connected to GC1000 Mark II>

5-43

IM 11B06C01-01E

CAUTION

If peak configuration or the total number of peak number assignments is modified while the

analysis results window is activated, re-establish individual connection with the analyzer to re-

acquire the database from the analyzer. Continuing to perform operations without re-establishing

individual connection with the analyzer may cause added or modified data to be improperly

displayed in the window.

• Data in the window can be converted to a graph using the Create command in the Graph

menu. (Only calibration coefficient data can be turned into a graph; no standard area data

can be made into a graph.)

l

Window configuration

The following shows the configuration of the Calibration Coefficient History window.

F5-4-7.ai

l

Display contents

The following lists the display contents for each item.

Item

Display Contents

#

Indicates the absolute numbers of peaks (serial numbers for all peaks of all

tasks): 1 to 255.

Stream # - Peak #

- Peak Name

Indicates all peaks that have been assigned numbers in the analyzer, in an

ascending order of stream numbers and of peak numbers (order of assignment).

A stream number, peak number, and peak name are connected and indicated in

this order using “-”, and the concentration unit (%, ppm, or no unit) of that peak is

indicated in parentheses at the end.

Date/Time

Indicates the date (year/month/day) and time (hour:minute) at which analysis

was started, for each analysis cycle (250 times maximum).

Calibration Coefficient

Indicates the calibration coefficient for each peak in the upper part of the Data

Display area.

Standard Area

Indicates the standard area for each peak in the lower part of the Data Display

area.

5th Edition : Nov. 16, 2011-00