<4. Display and Operations while Connected to GC8000>

4-14

IM 11B06C01-01E

4.3.6

Changing Calibration/Validation No.

To change a Calibration/Validation number, select the Calibration/Validation command from the

Measurement Status menu. The Calibration/Validation dialog box appears.

List only those for which calibration

settings have been specified.

Figure 4.17

Calibration/Validation dialog box

From the Calibration/Validation dialog box, you can select any of Calibration 1 to Calibration 6,

and any of Validation 1 to Validation 6.
If the stream number is not set, its option is not displayed. (If no stream number is set for

Calibration/Validation, its button is grayed out and cannot be selected from the dialog box.)
When you specify a Calibration/Validation number and click the Execute button, the

measurement status is changed to the Calibration/Validation of the selected calibration/validation

number.
If you select a Dummy Task (Stream 33), the Stream under Measurement is changed to the

“Replacement available.” In this case, the Dummy Task (Stream 33) is measured continuously

until you select the Start Calibration/Validation command from the Measurement Status menu.

(The chromatogram is not saved and the analysis result is not calculated.)

4.3.7

Changing Calibration/Validation Method

To change the Calibration/Validation Method, select the Calibration/Validation Method command

from the Measurement Status menu. The Calibration/Validation Method dialog box appears.

Figure 4.18

Calibration/Validation Method dialog box

5th Edition : Nov. 16, 2011-00