16 detector signal display, Detector signal display -19 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 66

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-19

IM 11B06C01-01E

4.3.16

Detector Signal Display

To display detector signals, right-click the detector. The Detector Signal dialog box appears.

Figure 4.23

Detector Signal dialog box (The TCD at left, FID-MC at center, and FPD (similar to FID) at

right)

The display is refreshed in every second. The following items are displayed.

Table Detector signal display item

Item

Description

Type

Indicates the detector type. It is either “TCD”, “FID”, “FID-MC” (FID having methane

converter), “FPD”, or “None”.

Input Signal

The input signal is sampled every 20 msec, and this analog signal is converted into

digital form, averaged at the sampling rate, and displayed.

Filtered Signal

The input signal is filtered using the filter constant of detector signal settings and

displayed.

Standard Deviation

Indicates the standard deviation of the past 20 input signals.

Voltage

Indicates the TCD bridge voltage. (TCD only)

Current

Indicates the TCD bridge current. (TCD only)

Flame Level

Indicates the preset Flame Detection Level. (FID, FID-MC, FPD)

Therm. Signal

Displays the thermocouple signals. (FID, FID-MC, FPD)

Flame Status

Indicates the Flame Detection status (Flame On or Flame Out). (FID, FID-MC, FPD)

Voltage of MC

Indicates the converter voltage. (FID-MC only)