<5. Display and Operations while Connected to GC1000 Mark II>

5-39

IM 11B06C01-01E

CAUTION

When an operation pattern or peak value setting is changed while an analysis results window is

being displayed:
(First close the analysis results window by executing the Exit command in the File menu, and

exit the Engineering Terminals. Then restart the Engineering Terminals and re-open the analysis

results window. Continuing operations without restarting all the Engineering Terminals may result

in incorrect data display.

l

Display contents

The following lists the display contents for each item.

Item

Display Contents

Analysis Start Time

Displays the date and time at which analysis of display data was started.

#

Indicates the absolute numbers of peaks (serial numbers for all peaks of all

tasks): 1 to 255.

Stream# - Peak# - Peak

Name

Indicates the peaks of a stream for which the latest analysis has been made, in

the ascending order of peak numbers (order of assignment).

A stream number, peak number, and peak name are connected and indicated in

this order using “-”, and the concentration unit (%, ppm, or no unit) of that peak is

indicated in parentheses at the end.

Concentration

Indicates concentration for each peak (with the unit indicated after the peak

name).

Retention Time

Indicates Retention time (second) for each peak.

Det. #

Indicates detector number (1 or 2) for each peak.

Peak

Peak start/Retention/

Peak end

Indicates peak process start time/Retention time/end time (second) for each

peak.

Gate

Gate start/Gate end

Indicates gate start time/end time (second) for each peak.

Pursuit

Indicates the peak pursuing status for each peak as “standard”, “pursuit”, or

“none”.

Peak Level

Start/Height/End

Displays the peak start level/top level/end level (mV) for each peak.

Area

Indicates the integral value of the peak level (mV second) for each peak.

Shape

Indicates the peak data process status in two-digit letters for each peak. It

displays “P” for tangent correction re-setting or “H” for the perpendicular method.

Half-value Width

Indicates a half-value width (second) for each peak.

Tailing Constant

Indicates a tailing constant for each peak.

Variation Coefficient

Indicates a concentration variation coefficient for each peak.

Method

Indicates a concentration computation method for each peak, selecting

the applicable concentration computation method from “corrected area,”

“calibration,” “external third,” “external linear,” “indirect method.”

(If no applicable method is found, “????????” is indicated.)

* Note: If commas (,) are used in stream names or peak names, the features of the analysis results windows may not function properly.

5th Edition : Nov. 16, 2011-00