<4. Display and Operations while Connected to GC8000>

4-6

IM 11B06C01-01E

4.2.3 User Level

In addition to the conventional user levels, level “C+” has been added for connection to the

GC8000. User level A has the lowest privilege, and user level C+ has the highest privilege. The

privileges assigned to each user level are summarized below.

Table Type of User Levels and Privileges

User Level

Privilege

Password

A

Allows the user to check the operation status and analysis

results only.

Does not allow the analyzer operation.

Not required (Default)

B

Among the analyzer operations, this level allows the user

to change the operation mode, change the measurement

status (except for cancel specification), and to change the

range.

1192

C

Allows the user to check and set up all items of analyzer

operation. However, re-integration of chromatogram and

editing of analysis results are not allowed.

1603

C+

Allows re-integration of chromatogram and editing of

analysis results.

1702

You can change the user level in the following procedure.

l

Changing the user level using the Change User Level button on the toolbar

1) Click the Change User Level button on the toolbar, and the Set User Level dialog box

appears.

2) From the Set User Level dialog box, select the desired user level. Then, click the OK button

and the Set Password dialog box appears.

3) Enter the password and click the Execute button. When the password is correct, a message

box showing “User level is set to xx” appears (where, “xx” is the new user level).

l

When changing the user level using the User Level command of Connect

menu

1) From the User Level command list of Connect menu, select the desired user level (A to C+).
2) The Password Input dialog box appears.
3) Enter the password and click the Execute button. When the password is correct, a message

box showing “User level is set to xx” appears (where, “xx” is the new user level).

Note: No password input is requested when you change to user level A.

Figure 4.5

Set User Level dialog box