94 • Start Up, Bar Hole Mode

Gas Tracer Operator’s Manual

The alarm LED’s and buzzer will pulse several times. After
this, press the RESET SILENCE button until you hear a beep to
continue.

If the unit is due for a bump test and BP EXPRD is set to NOT
USE, then the following screen displays.

The Gas Tracer cannot be used until a bump test has been
performed by selecting BUMP in the Maintenance Mode
menu. See “Performing a Bump Test” on page 129 for bump
testing instructions.

If a bump test is not due or BP EXPRD is set to NO EFFECT,
then the following screen appears for a few seconds indicating
when the next bump test is due.

NOTE:

If both CL RMNDR and BP RMNDR are turned off, a WARM
UP screen will display before the warm up sequence
continues.

FAIL


B--LIMIT

2011
9
30
10d

NEXTBUMP