Gas Tracer Operator’s Manual

Display Mode • 51

The table below illustrates how much data logging time is available
for the various interval times. It assumes that the unit is set up with all
5 sensors and there are no alarm occurrences.

Clear Data Logger Screens

CAUTION:

Once you clear the data logger, you cannot retrieve any data
previously stored in the data logger.

NOTE: This item appears in Display Mode only if the DLOG DSP menu

item in Maintenance Mode is turned ON. The factory setting for
DLOG DSP is ON.

The clear data logger screens allow you to clear the data logger
storage. You can set the Gas Tracer to overwrite the oldest data when
the data log is full. See “Updating the Datalog Overwrite Setting” on
page 160. The bottom line of the screen alternates between LG
CLEAR, NO /DISP, and YES/ENT.

To clear the data log:

1. With the above screen displayed, press and release the POWER

ENTER button. The instrument will ask you to confirm that you
want to clear the data log and CLr LOG will be flashing. The

Table 6: Data Logging Capacity, 5 Sensor Gas Tracer

Interval

Time

Data Logging Time

10 seconds

10 hours

20 seconds

20 hours

30 seconds

30 hours

1 minute

60 hours

3 minutes

180 hours

5 minutes

300 hours

10 minutes

600 hours

CLr


LG CLEAR

LOG