150 • Using Maintenance Mode

Gas Tracer Operator’s Manual

2. Press and release the POWER ENTER button. The screen will

indicate the current setting.

3. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

4. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Alarm Silence Setting

With ALM SLNC set to ON (factory setting), pressing and releasing
the RESET SILENCE button silences the buzzer when the Gas Tracer
is in alarm. The LEDs continue to flash and the display continues to
show the alarm. If the alarms are set up as latching, when the gas
concentration falls below the alarm level, pressing and releasing
RESET SILENCE button clears all alarm indications for that alarm. If
an alarm condition occurs, you may still enter Display Mode by
pressing the DISPLAY (ADJ) button. The buzzer will be silenced but
the LEDs will continue to flash and the unit will continue to vibrate. If
you return to Measuring Mode and there is still an alarm condition,
the LEDs will continue to flash, the unit will continue to vibrate, and
the buzzer will remain off. Once the condition clears, press the RESET
SILENCE button to clear the alarm indications.

With ALM SLNC set to OFF, you cannot silence the buzzer. If an
alarm condition occurs, and you enter Display Mode, the buzzer will
not be silenced and the LEDs will continue to flash. Upon return to
Measuring Mode, if there is still an alarm condition and the alarms are
set up as latching, you must wait until it clears before you can press
the RESET SILENCE button to clear the alarm indications.

On


MAINTENANCE

LATCHING