Co with h, S response configuration – RKI Instruments Gas Tracer User Manual

Page 42

Advertising
background image

34 • Measuring Mode, Normal Operation

Gas Tracer Operator’s Manual

you are detecting pentane and the display reads 10% LEL on the combustible
gas channel, you actually have 10% LEL x 1.35 = 13.5% LEL pentane
present.

CO with H

2

S Response Configuration

Some applications may require the user to know if H

2

S is present

while monitoring with a Gas Tracer. The Gas Tracer does not have an
H

2

S sensor installed but if the charcoal filter is removed from the CO

sensor, it will respond to H

2

S.

Since the CO sensor cannot distinguish between CO and H

2

S, the CO

channel reading on the display could be a mix of CO and H

2

S.

If the environment you are monitoring is completely free of CO, there
is an approximate response factor that varies from sensor to sensor
that can be used. The CO sensor’s response to H

2

S is 5 times greater

than the actual H

2

S concentration. For example, if the CO channel

reading is 25 ppm, the actual H

2

S concentration is 5 ppm. This

approximation is not applicable if CO is present in the monitoring
environment.

NOTE:

If your Gas Tracer is configured to respond to both H

2

S and

CO, be sure to calibrate the CO channel using CO.

Table 3: LEL Hydrocarbon Conversions

Gas

LEL Conversion

Factor (Methane

Calibration)

Gas

LEL Conversion

Factor (Methane

Calibration)

Acetone

1.92

Iso Butane

1.56

Acetylene

2.00

MEK

1.92

Benzene

2.00

Methane

1.00

Ethane

1.25

Methanol

1.65

Ethanol

1.75

Pentane

1.35

Ethylene

1.20

Propane

1.52

Heptane

1.92

Propylene

1.33

Hexane

1.65

Toluene

2.00

Hydrogen

1.00

Xylene

0.93

IPA

2.50

Advertising