72 • Calibration Mode

Gas Tracer Operator’s Manual

3. Install the hose and probe to the Gas Tracer.

4. Use the AIR▲ or (SHIFT)▼ buttons to display the BUMP menu

item.

5. Press and release the POWER ENTER button to display the bump

test gas values.

6. Use the AIR▲ or (SHIFT)▼ buttons to see the %volume

combustible gas bump test value.

7. A bump test may be performed on the standard 3 sensors or on the

%volume combustible sensor. Use the AIR▲ or (SHIFT)▼ buttons
to display the screen of sensor(s) you wish to bump test.

8. Before proceeding, confirm that the bump test gas value(s) are the

same as the concentration(s) in the calibration cylinder. If they are
not, adjust the bump test gas value(s) by entering the AUTO CAL
menu item, changing the values there, and reentering the BUMP
menu.

9. Connect the tubing from the demand flow regulator to the rigid

tube on the probe then quickly press and release the POWER


MAINTENANCE

BUMP

50
12.0
50

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

BUMP 30

CO

100


vol%

MAINTENANCE

CH

4

BUMP 30