Bp rmndr, Updating the bump test reminder setting, Updating the bump test expired action setting – RKI Instruments Gas Tracer User Manual

Page 173

Advertising
background image

Gas Tracer Operator’s Manual

Using Maintenance Mode • 165

Updating the Bump Test Reminder Setting

With BP RMNDR set to ON, the Gas Tracer will give an indication at
start up if it is due for bump testing. The type of indication will
depend on the BP EXPRD setting (see below).

With BP RMNDR set to OFF (factory setting), the Gas Tracer will give
no indication at start up if it is due for bump testing.

NOTE:

If the BUMP DSP menu item is set to OFF, this menu item
does not appear.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the BP RMNDR

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Bump Test Expired Action Setting

This item defines what indication is given during start up when a
bump test is due and BP RMNDR is set to ON.

With BP EXPRD set to CONFIRM (factory setting), the Gas Tracer
will give an indication at start up if a bump test is due and require the
user to press the RESET SILENCE button to continue.

With BP EXPRD set to NOT USE, if the unit is due for a bump test the
Gas Tracer will give an indication at start up that the bump test is due
and the unit cannot be used until it is bump tested. With this setting, it
is necessary to bump test the instrument from the Maintenance Mode
if a bump test is not performed before it is due. See “Performing a
Bump Test” on page 129 and “Calibrating and Bump Testing the PPM
Combustible Sensor” on page 135 for instructions.

With BP EXPRD set to NO EFFECT, no confirmation during startup
or bump test is required to use the Gas Tracer.

On


MAINTENANCE

BP RMNDR