Bp intvl, Updating the bump test interval – RKI Instruments Gas Tracer User Manual
Page 172

164 • Using Maintenance Mode
Gas Tracer Operator’s Manual
Updating the Bump Test Interval
The BP INTVL setting defines the amount of time between bump
tests. The minimum setting is 0 days and the maximum setting is 30
days. The factory setting is 30 days.
NOTE:
If the BUMP DSP menu item is set to OFF, this menu item
does not appear.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BP INTVL
menu item. Press and release the POWER ENTER button. The
screen indicates the current setting.
2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.
Tip:
Press and hold the AIR▲ or (SHIFT)▼ button to rapidly scroll
through the settings.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
30
MAINTENANCE
BP INTVL