152 • Using Maintenance Mode

Gas Tracer Operator’s Manual

2. Press and release the POWER ENTER button. The screen will

indicate the current setting.

3. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

4. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Alarm Point Settings

The ALARM--P menu item allows you to update one or more
warning, alarm, STEL, and TWA (for CO) settings (the reading at
which the Gas Tracer recognizes these conditions).

For the combustible gas channel, it is not possible to adjust either the
Warning or Alarm settings to be higher than 60 %LEL. In addition, the
Warning setting cannot be adjusted higher than the Alarm setting for
all channels.

NOTE:

It is not necessary to adjust every setting for every channel
when you enter this menu item. All settings are shown here
for example.

Table 7: Factory Set Alarm Points

Channel

Warning

Alarm

STEL

TWA

Combustible Gas

10% LEL

50% LEL

N/A

N/A

Oxygen

19.5% decreasing

23.5% increasing

N/A

N/A

CO

25 ppm

50 ppm

200 ppm

25 ppm

300