40 • Measuring Mode, Alarms

Gas Tracer Operator’s Manual

Resetting and Silencing Alarms

You can set the Gas Tracer’s gas alarms as latching or self-resetting
alarms (See “Updating the Alarm Latching Setting” on page 149) and
the buzzer operation as silenceable or not silenceable (See “Updating
the Alarm Silence Setting” on page 150).

• Self-resetting alarms (LATCHING set to OFF)

Self-resetting alarms automatically shut off and reset when the gas
reading falls below (or rises above for an oxygen low alarm) the
alarm setting. You cannot silence or reset self-resetting alarms.

• Latching alarms (LATCHING set to ON)

You can set latching alarms with or without Alarm Silence (see
“Updating the Alarm Silence Setting” on page 150).

With ALRM SLNC On and LATCHING On:

When the Gas Tracer goes into a gas alarm, press the RESET SILENCE
button to silence the buzzer. If the gas concentration was still above
the alarm level when the button was pressed, the LEDs continue to

Sensor Failure*

FAIL appears at the

bottom of the display
and the failed
sensor(s) are
indicated

• Alarm LED arrays

flash

Buzzer sounds a double
pulsing tone (two pulses
in quick succession)

Clock Failure*

• FAIL CLOCK appears

on the display

• Alarm LED arrays

flash

Buzzer sounds a double
pulsing tone (two pulses
in quick succession)

System Failure*

• FAIL SYSTEM

appears on the
display

• Alarm LED arrays

flash

• Buzzer sounds a

double pulsing tone
(two pulses in quick
succession)

• A failure code appears

Microprocessor
Failure*

• NOTE: The unit will

not operate if this
alarm occurs.

• Heart indicator is

steadily on or not on
at all

• None

*

This alarm can also occur in Bar Hole Mode

Table 4: Alarm Types and Indications

Alarm Type

Visual Indications

Other Indications