Bump 30 – RKI Instruments Gas Tracer User Manual

Page 214

Advertising
background image

206 • Bump Testing

Gas Tracer Operator’s Manual

3. Press and release the POWER ENTER button to display the bump

test gas values.

4. Use the AIR▲ or (SHIFT)▼ buttons to see the %volume

combustible gas bump test value.

5. A bump test may be performed on the standard 3 sensors or on the

%volume combustible sensor. Use the AIR▲ or (SHIFT)▼ buttons
to display the screen for the sensor(s) you wish to bump test.

6. Before proceeding, confirm that the bump test gas value(s) are the

same as the concentration(s) in the calibration cylinder. If they are
not, adjust the bump test gas value(s) by entering the AUTO CAL
menu item, changing the values there, and reentering the BUMP
menu.

7. Open the hose clamp and connect the tubing from the filled gas

bag to the rigid tube on the probe and quickly press and release the
POWER ENTER button. The unit will begin counting down from
30.

50
12.0
50

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

BUMP 30

CO

100


vol%

MAINTENANCE

CH

4

BUMP 30

Advertising