158 • Using Maintenance Mode
Gas Tracer Operator’s Manual
1. Use the AIRS or (SHIFT)T buttons to display the POFF DSP
menu item.
2. Press and release the POWER ENTER button. The screen will
indicate the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Bump Test Parameters
The BUMP--SET menu item is used to adjust the bump test
parameters.
NOTE:
If the BUMP DSP menu item is set to OFF, this menu item
does not appear.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BUMP--SET
menu item.
2. Press and release the POWER ENTER button. The values displayed
are for GAS TIME, CHECK (bump test tolerance), CAL TIME, and
MAINTENANCE
POFF DSP
OFF
MAINTENANCE
POFF DSP
MAINTENANCE
BUMP--SET