Cl rmndr, Updating the calibration reminder setting, Updating the calibration expired action setting – RKI Instruments Gas Tracer User Manual

Page 170

Advertising
background image

162 • Using Maintenance Mode

Gas Tracer Operator’s Manual

Updating the Calibration Reminder Setting

With CL RMNDR set to ON (factory setting), the Gas Tracer will give
an indication at start up if it is due for calibration. The type of
indication will depend on the CL EXPRD setting (see below).

1. Use the AIR▲ or (SHIFT)▼ buttons to display the CL RMNDR

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Calibration Expired Action Setting

This item defines what indication is given during start up when
calibration is due and CL RMNDR is set to ON.

With CL EXPRD set to CONFIRM (factory setting), the Gas Tracer
will give an indication at start up if calibration is due and require the
user to press the RESET SILENCE button to continue.

With CL EXPRD set to NOT USE, if the unit is due for calibration the
Gas Tracer will give an indication at start up that calibration is due
and the unit cannot be used until it is calibrated. With this setting, it is
necessary to calibrate the instrument from Maintenance Mode if
calibration is not performed before it is due. See “Calibrating Using
Auto Calibration” on page 123, “Calibrating Using Single
Calibration” on page 126, or “Calibrating and Bump Testing the PPM
Combustible Sensor” on page 135 for instructions.

With CL EXPRD set to NO EFFECT, no confirmation during startup
or calibration is required to use the Gas Tracer if calibration is
overdue.