88 • Calibration Mode

Gas Tracer Operator’s Manual

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

10.When AUTO CAL in the BUMP--SET menu item in Maintenance

Mode is set to ON:

If the ppm combustible sensor passes the bump test,

The following screen appears:

To view the bump test gas reading press the AIR▲ or
(SHIFT)▼ button. You can scroll between the bump test results
and the bump test gas reading with the AIR▲ or (SHIFT)▼
button.

To return to the BUMP screen in PPM CAL, press POWER
ENTER.

Disconnect the humidifier from the probe.

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

P


ppm

MAINTENANCE

CH

4

BUMP

1000


ppm

MAINTENANCE

CH

4

BUMP


MAINTENANCE

BUMP