Gas Tracer Operator’s Manual

Calibration Mode • 75

lower right of the screen. The calibration time counted down
during a calibration initiated because of a failed bump test is
the difference between the bump test gas time and the
calibration time defined in the BUMP--SET menu item in
Maintenance Mode.

At the end of the calibration, the instrument displays the
results from both the bump test and the calibration. Use the
AIR▲ or (SHIFT)▼ buttons to scroll between the calibration/
bump test results, the bump test gas readings, and the
calibration gas readings.

1
20.9
37

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

CAL 60

CO

FP
PP
FF

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

BUMP/CAL

CO

1
12.0
5

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

BUMP/

CO