204 • Bump Testing

Gas Tracer Operator’s Manual

11.After the last channel is calibrated, disconnect the calibration

tubing from the probe.

12.Use the AIR▲ and (SHIFT)▼ buttons to scroll to ESCAPE in the

ONE CAL menu.

13.Press and release the POWER ENTER button to exit ONE CAL.

14.With the ONE CAL menu option displayed, press the (SHIFT)▼

button until the NORMAL

menu option is displayed.

15.Press and release the POWER ENTER button to return to

Measuring Mode.

Bump Testing

NOTE:

The BUMP DSP menu item in Maintenance Mode is factory
set to OFF. The BUMP menu item will not appear unless
BUMP DSP

is set to ON. See “Turning the Bump Test

Function On or Off” on page 157 for instructions.

NOTE:

The Gas Tracer uses parameters defined in the BUMP--SET
menu item of Maintenance Mode while performing a bump
test. To view or update these parameters, see “Updating the
Bump Test Parameters” on page 158.

NOTE:

If the combustible gas channels are calibrated to a gas other
than methane, use an appropriate multigas cylinder.