198 • Calibrating the Gas Tracer

Gas Tracer Operator’s Manual

you are returned to the initial AUTO CAL screen and must begin
the auto calibration process again.

2. Fill the gas bag with calibration gas as described in “Filling the Gas

Bag” on page 195.

3. Use the AIR▲ and (SHIFT)▼ buttons to scroll to the AUTO CAL

menu item. Press and release the POWER ENTER button to display
the calibration values screen. If you want to exit back to the main
menu without completing a calibration, press and release the
DISPLAY (ADJ) button.

4. Use the AIR▲ or (SHIFT)▼ buttons to scroll between the 3-gas and

the %volume combustible gas calibration values screens.

5. The gas concentrations displayed in the calibration values screens

must match the gas concentrations listed on the calibration
cylinders. If all concentrations match, go to step 13. If one or more
concentrations do not match, continue with step 6.