138 • Using Maintenance Mode

Gas Tracer Operator’s Manual

1. Once the warm up is complete, you will be in PPM Cal Mode and

AIR CAL will be displayed. You must perform a fresh air adjust
before performing an auto calibration, a single calibration, or a
bump test.

2. Press and release the POWER ENTER button to enter the AIR CAL

menu item.

You will be prompted to choose between WET and DRY.

3. Use the AIR▲ or (SHIFT)▼ buttons to display the DRY setting and

press and release the POWER ENTER button. The unit will display
the current fresh air reading and prompt you to push the AIR▲
button. Do not press the AIRbutton at this time.

4. Install the demand flow regulator onto the zero air cylinder.

5. Connect the sample tubing to the demand flow regulator.

6. Install the hose and probe to the Gas Tracer.

7. Connect the tubing from the demand flow regulator to the rigid

tube on the probe. Allow the zero air to flow for two minutes.


MAINTENANCE

AIR CAL


MAINTENANCE

WET

50


ppm

MAINTENANCE

CH

4

PUSH AIR