128 • Using Maintenance Mode

Gas Tracer Operator’s Manual

to calibrate (in this example the combustible gas % LEL sensor). If
you have a 5 sensor unit, the %vol sensor will be one of the choices.
It is shown below.

8. Press and release the POWER ENTER button. The single

calibration screen displays for the sensor you selected. The gas
reading flashes.

9. Connect the tubing from the demand flow regulator to the rigid

tube on the probe. Allow the calibration gas to flow for one minute.

10.If necessary, use the AIR▲ and (SHIFT)▼ buttons to adjust the

reading to match the concentration listed on the calibration
cylinder.

If the reading already matches the concentration listed on the
calibration cylinder, press and release the AIR▲ button then press
and release the (SHIFT)▼ button so that the instrument sees a
change and resets the calibration date.

11.Press and release the POWER ENTER button to save the span

value. The LCD will indicate that the calibration has ended, then
the single calibration menu displays.

- - -