Gas Tracer Operator’s ManualCalibrating and Bump Testing the PPM Combustible Sensor

bump test.

2. If the instrument is not in a fresh air environment, fill the ga bag

with zero air as described in “Filling the Gas Bag” on page 195.

3. Press and release the POWER ENTER button to enter the AIR CAL

menu item.

NOTE:

If you have just performed a DRY fresh air adjustment, allow
the Gas Tracer to sample fresh air for 2 minutes before
continuing.

You will be prompted to choose between WET and DRY.

4. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The unit will display
the current fresh air reading and prompt you to push the AIR▲
button.

5. Be sure the Gas Tracer is in a fresh air environment. If the

instrument is not in a fresh air environment, connect a gas bag with
zero air to the rigid tube on the probe using the 6 inch humidifier


MAINTENANCE

AIR CAL


MAINTENANCE

WET

50


ppm

MAINTENANCE

CH

4

PUSH AIR