Gas Tracer Operator’s Manual

Start Up, Leak Check Mode • 103

6. If CL RMNDR is turned on (factory setting), the screen that

appears next depends on how CL EXPRD is set in the Maintenance
Mode Menu (page 162).

If the unit is due for calibration and CL EXPRD is set to
CONFIRM (factory setting), then the following screen displays.

The alarm LED’s and buzzer will pulse several times. After
this, press and release the RESET SILENCE button to continue.

If the unit is due for calibration and CL EXPRD is set to NOT
USE, then the following screen displays.

The Gas Tracer cannot be used until a calibration has been
performed either by selecting AUTO CAL, ONE CAL, or PPM
CAL

in the Maintenance Mode menu. See “Calibrating Using

Auto Calibration” on page 123, “Calibrating Using Single
Calibration” on page 126, or “Calibrating and Bump Testing
the PPM Combustible Sensor” on page 135 for calibration
instructions.

CAL


C--LIMIT

FAIL


C--LIMIT