Gas Tracer Operator’s Manual
Using Maintenance Mode • 155
11.Adjust the values for the warning and alarm settings using the
procedure in step 4 through step 7. When you press POWER
ENTER after selecting the alarm value, you will then be prompted
to select the STEL value.
12.Use the AIR▲ or (SHIFT)▼ buttons to change the setting and press
POWER ENTER when finished to save the setting. You will then be
asked to enter a TWA value.
13.Use the AIR▲ or (SHIFT)▼ buttons to change the setting and press
POWER ENTER when finished to save the setting and return to the
Alarm Points Screen.
14.When you are finished making adjustments, use the AIR▲ or
(SHIFT)▼ buttons to select ESCAPE and press POWER ENTER.
You will be returned to the ALARM--P screen in the main menu.
200
ppm
MAINTENANCE
STEL
CO
25
ppm
MAINTENANCE
TWA
CO
- - -
ppm
MAINTENANCE
ALARM--P
CO