Gas Tracer Operator’s Manual

Using Maintenance Mode • 155

11.Adjust the values for the warning and alarm settings using the

procedure in step 4 through step 7. When you press POWER
ENTER after selecting the alarm value, you will then be prompted
to select the STEL value.

12.Use the AIR▲ or (SHIFT)▼ buttons to change the setting and press

POWER ENTER when finished to save the setting. You will then be
asked to enter a TWA value.

13.Use the AIR▲ or (SHIFT)▼ buttons to change the setting and press

POWER ENTER when finished to save the setting and return to the
Alarm Points Screen.

14.When you are finished making adjustments, use the AIR▲ or

(SHIFT)▼ buttons to select ESCAPE and press POWER ENTER.
You will be returned to the ALARM--P screen in the main menu.



200

ppm

MAINTENANCE

STEL

CO



25

ppm

MAINTENANCE

TWA

CO



- - -

ppm

MAINTENANCE

ALARM--P