112 • Leak Testing

Gas Tracer Operator’s Manual

second beep and the Leak Check Mode Screen displays.

Snap Log Mode, Leak Check Mode Operation

The snap logging function in Snap Log Mode allows the user to record
data at a specific time and have it saved to the data logger. The data is
assigned a snap log ID and is saved with the station ID that was
selected when the data was taken.

NOTE:

If CO DISP in Maintenance Mode is turned on and a CO
reading is displayed, that CO reading will not be saved in the
snap log data.

To enter Snap Log Mode and record snap log data:

1. Turn the Gas Tracer on as described in “Turning On the Gas Tracer,

Leak Check Mode” on page 101. Select the desired display range.
In the example below, 500 ppm has been selected. Press and hold
the (SHIFT)T button, then press and hold the AIRS button and
hold both until you hear a beep. The Station ID Select Screen will
display and the current station ID will be flashing.

NOTE:

If the gas levels reach a warning or alarm condition while in
this screen, the Gas Tracer will go into alarm.

S Id