F p f, Bump – RKI Instruments Gas Tracer User Manual
Page 215

Gas Tracer Operator’s Manual
Bump Testing • 207
8. When AUTO CAL in the BUMP--SET menu item in Maintenance
Mode is set to OFF:
•
The instrument will display which channels passed or failed
the bump test with a P (pass) or an F (fail) to the right of the
gas.
•
To view the bump test gas readings press the AIR▲ or
(SHIFT)▼ button. You can scroll between the bump test results
and the bump test gas readings with the AIR▲ or (SHIFT)▼
button.
•
To return to the BUMP screen in Calibration Mode, press
POWER ENTER.
See “Troubleshooting” on page 175 to investigate the cause of the
failure and replace the failed sensor or sensors if necessary.
•
Disconnect the tubing from the probe.
F
P
F
%LEL
%
ppm
MAINTENANCE
CH
4
O
2
BUMP
CO
1
12.0
5
%LEL
%
ppm
MAINTENANCE
CH
4
O
2
BUMP
CO
MAINTENANCE
BUMP