One cal – RKI Instruments Gas Tracer User Manual

Page 210

Advertising
background image

202 • Calibrating the Gas Tracer

Gas Tracer Operator’s Manual

1. While in the calibration menu, use the AIR▲ or (SHIFT) ▼ button

to scroll to the ONE CAL menu option.

2. Fill the gas bag with calibration gas as described in “Filling the Gas

Bag” on page 195.

3. Press and release the POWER ENTER button to display the single

calibration menu. The %LEL sensor will be displayed first.

4. Use the AIR▲ or (SHIFT)▼ button to display the sensor you want

to calibrate (in this example the combustible gas % LEL sensor). If
you have a 5 sensor unit, the %vol sensor will be one of the choices.
It is shown below.


MAINTENANCE

ONE CAL

- - -


%LEL

MAINTENANCE

CH

4

ONE CAL

- - -


vol%

MAINTENANCE

CH

4

ONE CAL