One cal – RKI Instruments Gas Tracer User Manual

Page 150

Advertising
background image

142 • Using Maintenance Mode

Gas Tracer Operator’s Manual

5. You will be prompted to choose between WET and DRY. Use the

WET setting.

NOTE:

The DRY setting in ONE CAL is for factory use only. Do not
perform a calibration in ONE CAL using the DRY setting.

6. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The gas reading will
be replaced by 3 dashes.

7. Press and release the POWER ENTER button. The single

calibration screen displays. The gas reading flashes.

8. Connect the humidifier from the demand flow regulator to the

rigid tube on the probe. Allow the calibration gas to flow for two
minutes.

9. If necessary, use the AIR▲ and (SHIFT)▼ buttons to adjust the

reading to match the concentration listed on the calibration
cylinder.

If the reading already matches the concentration listed on the
calibration cylinder, press and release the AIR▲ button then press
and release the (SHIFT)▼ button so that the instrument sees a
change and resets the calibration date.

10.Press and release the POWER ENTER button to save the span

value. The LCD will indicate that the calibration has ended, then
the single calibration menu displays.

11.Disconnect the humidifier from the probe.

- - -


ppm

MAINTENANCE

CH

4

ONE CAL

0


ppm

MAINTENANCE

CH

4

ONE CAL