Gas Tracer Operator’s ManualCalibrating and Bump Testing the PPM Combustible Sensor

To return to the BUMP screen in PPM CAL, press POWER
ENTER at any time. If the sensor fails calibration, the buzzer
will sound a double pulsing tone until you press and release
the POWER ENTER button.

Disconnect the humidifier from the probe.

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

9. Use the (SHIFT)▼ button to navigate to the ESCAPE menu item,

then press and release the POWER ENTER button to return to the
PPM CAL

menu item in Calibration Mode.

10.Use the (SHIFT)▼ button to navigate to the NORMAL menu item,

then press and release the POWER ENTER button to return to
Measuring Mode.