Specifications – RKI Instruments Gas Tracer User Manual

Page 11

Advertising
background image

Gas Tracer Operator’s Manual

Specifications • 3

Specifications

Table 1: Gas Tracer Specifications

Target Gas

• %LEL

Combustible Gas
(Methane
Calibration
Standard)*

• PPM Combustible

Gas (Methane
Calibration
Standard)*

% Volume

Combustible

Gas

(Methane

Calibration

Standard)

Oxygen (O

2

)

Carbon

Monoxide

(CO)

Range

(Increment)

0-100% LEL

(1% LEL)

0 - 100% vol

(1% vol)

0-40.0% vol

(0.1 vol%)

0-500 ppm

(1 ppm)

Leak Check

Mode Range

(Increment)

0-5000 ppm

(10 ppm)

NOTE: Detection
range is 0-5000
ppm, but displayed
range can be 0-500,
0-1000, 0-2000 or
0-5000 ppm.

n/a

n/a

n/a

Sampling Method

Sample Draw

Response Time

T90 Within 30 Seconds

Accuracy

LEL: ± 5% of reading

or ± 2% LEL

(whichever is

greater)

PPM: ± 30 ppm or ±

5% of reading

(whichever is

greater)

± 5% of

reading or ±

2% full scale

(whichever

is greater)

± 0.5% O

2

± 5% of

reading or ±

5 ppm CO

(whichever is

greater)

Display Digital

LCD

Display

Gas Alarms

(Factory Settings)

Alarm 1
10% LEL

Alarm 2
50% LEL

None

Alarm 1
19.5 vol%
(Decreasing)

Alarm 2
23.5 vol%
(Increasing)

Alarm 1
25 ppm
Alarm 2
50 ppm
TWA
25 ppm
STEL
200 ppm

Operating

Temperature &

Humidity

-20°C to 50

°

C/Below 85% RH (Without Condensation)

Advertising