74 • Calibration Mode

Gas Tracer Operator’s Manual

11.When AUTO CAL in the BUMP--SET menu item in Maintenance

Mode is set to ON:

If all channels pass the bump test,

The following screen appears:

To view the bump test gas readings press the AIR▲ or
(SHIFT)▼ button. You can scroll between the bump test results
and the bump test gas readings with the AIR▲ or (SHIFT)▼
button.

To return to the BUMP screen in Calibration Mode, press
POWER ENTER.

Disconnect the tubing from the probe.

If any channel fails the bump test,

The instrument will beep and a calibration is immediately and
automatically started. Continue to apply the calibration gas.
CAL on the bottom left of the screen will alternate with APPLY.

The calibration time will begin to countdown in seconds in the