Bump, Cal 60, Fp pp ff – RKI Instruments Gas Tracer User Manual

Page 141: Bump/cal

Advertising
background image

Gas Tracer Operator’s Manual

Using Maintenance Mode • 133

To return to the main menu, press POWER ENTER.

Disconnect the tubing from the probe.

If any channel fails the bump test,

The instrument will beep and a calibration is immediately and
automatically started. Continue to apply the calibration gas.
CAL on the bottom left of the screen will alternate with APPLY.

The calibration time will begin to countdown in seconds in the
lower right of the screen. The calibration time counted down
during a calibration initiated because of a failed bump test is
the difference between the bump test gas time and the
calibration time defined in the BUMP--SET menu item in
Maintenance Mode.

At the end of the calibration, the instrument displays the
results from both the bump test and the calibration. Use the
AIR▲ or (SHIFT)▼ buttons to scroll between the calibration/
bump test results, the bump test gas readings, and the
calibration gas reading.


MAINTENANCE

BUMP

1
20.9
37

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

CAL 60

CO

FP
PP
FF

%LEL

%

ppm

MAINTENANCE

CH

4

O

2

BUMP/CAL

CO

Advertising