134 • Using Maintenance Mode
Gas Tracer Operator’s Manual
•
To return to the BUMP screen in Maintenance Mode, press
POWER ENTER at any time. If any channel failed the
calibration, the buzzer will sound a double pulsing tone until
you press and release the POWER ENTER button.
•
Disconnect the tubing from the Gas Tracer’s probe.
13.Unscrew the demand flow regulator from the calibration cylinder.
14.With the BUMP menu option displayed, press the (SHIFT)▼
button until the NORMAL
menu option is displayed.
15.Press and release the POWER ENTER button to return to
Measuring Mode.