L./b. mode, Updating the leak check/bar hole mode setting – RKI Instruments Gas Tracer User Manual

Page 177

Advertising
background image

Gas Tracer Operator’s Manual

Using Maintenance Mode • 169

2. Use the AIR▲ or (SHIFT) ▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Leak Check/Bar Hole Mode Setting

With L./B. MODE turned OFF, the Gas Tracer will automatically enter
Normal Mode when it is turned on.

With L./B. MODE set to bH, you will be prompted to choose between
Normal Mode and Bar Hole Mode when the Gas Tracer is turned on.

With L./B. MODE set to LC (factory setting), you will be prompted to
choose between Normal Mode and Leak Check Mode when the
Gas Tracer is turned on.

With L./B. MODE set to LC bH, you will be prompted to choose
between Normal Mode, Bar Hole Mode, and Leak Check Mode when
the Gas Tracer is turned on.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the L./B. MODE

menu item.

2. Press and release the POWER ENTER button. The screen will

indicate the current setting.

3. Use the AIR▲ or (SHIFT) ▼ button to display the desired setting.

4. Press and release the POWER ENTER button to save the setting

and return to the main menu.




MAINTENANCE

L./B. MODE

OFF


MAINTENANCE

L./B. MODE

Advertising