Gas Tracer Operator’s Manual

Using Maintenance Mode • 163

1. Use the AIR▲ or (SHIFT)▼ buttons to display the CL EXPRD

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT) ▼ button to display the desired setting

along the bottom of the screen.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Calibration Check Setting

This item setting defines which sensors are used to update the
calibration reminder screen. C--CHECK can be set to 4-GAS or ALL. If
C--CHECK

is set to 4-GAS (factory setting), the calibration dates for

only the standard sensors (LEL, O

2

, and CO) are used to determine if

calibration is due. If C--CHECK is set to ALL, the calibration dates for
the standard sensors along with the %volume sensor and the ppm
combustible gas sensor are used to determine if a calibration is due.

1. Use the AIRS or (SHIFT)T buttons to display the C--CHECK

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIRS or (SHIFT)T button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.