5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
990
■■■■■■■■
tiamo 2.1
Stop volume
Volume of auxiliary solution (e.g. buffer) that is to be added to the sample
before the first standard addition.
Range
0.0 ... 9999.9 mL
Default value
100.0 mL
Addition solution
Note
Displayed only for Solution = not defined.
Concentration
Concentration of addition solution.
Range
0.001 ... 999999.999
Default value
1.000
Device
Concentration unit of addition solution.
Selection
mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L
Default value
ppm
Control parameters
Dosing rate
Speed at which the volume increments are added.
Selection
slow | medium | fast
Default value
fast
Delta U
Potential difference that is to be achieved by a standard addition.
Range
1 ... 999 mV
Default value
10 mV
5.6.4.16.4
STDADD auto - Measuring parameters
Tab: Method
▶ STDADD ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input