2 meas u - general/hardware, General/hardware – Metrohm tiamo 2.1 Manual User Manual

Page 888

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

874

■■■■■■■■

tiamo 2.1

Identification

Description

.IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value x (1…3) for the last measuring point
on the measuring point list

.LP.EXx

External value x (1…3) for the last measuring point on
the measuring point list

.LP.MEA

Measured value for the last measuring point in the
measuring point list in the unit of the measured value

.LP.TEM

Temperature for the last measuring point of the meas-
uring point list in °C

.LP.TIM

Time in s until the last measuring point in the measuring
point list is reached

.MA.MEA

Maximum measured value (U

ind

) in mV

.MA.TEM

Temperature for the maximum measured value in °C

.MA.TIM

Time in s until the maximum measured value is reached

.MI.MEA

Minimum measured value (U

ind

) in mV

.MI.TEM

Temperature with minimum measured value in °C

.MI.TIM

Time for the minimum measured value in s

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)

.NMP

Number of measuring points in the measuring point list

.SLO

Electrode slope of the sensor used for the command (in
mV for ISE sensors)

.STY

Type of stop with which the command was stopped:
1 = normal;0 = manual or after error

5.6.4.3.2

MEAS U - General/Hardware

Tab: Method

MEAS U Properties... General/Hardware

Command name

Name of the command.

Input

25 characters

Advertising