Metrohm tiamo 2.1 Manual User Manual
Page 1068

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1054
■■■■■■■■
tiamo 2.1
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The dosing rate should be reduced for viscous liquids.
Dosing time
Time during which dosing is to be carried out. This parameter is only dis-
played for Dosing criterion = Volume/Dosing time or Dosing rate/
Dosing time.
Range
0 ... 9999999 s
Default value
100 s
Pause
Waiting period before dosing is started.
Range
0 ... 999999 s
Default value
0 s
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously. This value is used for tem-
perature correction during monitoring with pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.