5 Method

tiamo 2.1

■■■■■■■■

837

Measured value drift

The measured value is only accepted if the drift is less than the value entered
here.

Range

0.1 ... 999.0 mV/min

Selection

off

Default value

off

off
The measured vales will not be applied until after the maximum waiting
time has passed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s

Pause 1

Pause 1

Waiting time, e.g. until the electrode has settled down, before a start volume
is added.

Titrando, 855
Range

0 ... 999999 s

Default value

0 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate indi-
cated.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL