5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

748

■■■■■■■■

tiamo 2.1

Parameters
The parameters for the command KFT Upol are configured in the following
8 tabs:

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titration.

Control parameters
Control parameters for the endpoint (measured value for the endpoint,
titration rate, user-defined parameters, stop criterion).

Titration parameters
Parameters for the run of the titration.

Stop conditions
Definitions of conditions which cause the stop of the titration.

Conditioning
Enable/Disable the conditioning. Definition of the start drift, drift correc-
tion and stop conditions.

Additional evaluations
Definition of additional methods for the evaluation of the titration curves.

Additional measured values
Definition of additional measured values of other measuring commands
which are saved as additional columns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command KFT Upol and can be used in formulas under the designation
'Command name.Variable designation':

Identification

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated

.COK

Command status; 1 = Condition requirement ful-
filled
, 0 = Condition requirement not fulfilled

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used
for the command

.DBL

Total duration for the processing of the command in s