5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1150

■■■■■■■■

tiamo 2.1

5.6.10.2.3

REQUEST - Sample data request

Dialog window: Workplace

[START] Sample data request

Message

Displaying the message defined in the command REQUEST.

Sample position

Position of the sample on the Sample Processor rack. This number can be
used for moving to a sample position with the command MOVE when
Target position = Sample position.

Range

1 ... 999

Default value

1

ID1 … ID16

Sample identifications.

Type = Number
Range

-1.0E-99 ... 1.0E+99

Type = Text
Input

max. 100 characters

Type = Date/Time
Selection

'YYYY-MM-DD' or 'YYYY-MM-DD hh:mm:ss'

Note

If the sample identifications are still invalid at the start of the method and
are not entered until the time of the REQUEST command during the run,
then the option Check at start in the START command must be
switched off for these method variables.

Sample size

Sample size (initial weight).

Input

1 ... 10 digits.

Default value

1

Device

Unit of sample size.

Selection

g | mg | µg | mL | µL | pcs

Default value

g