5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
934
■■■■■■■■
tiamo 2.1
Range
0 ... 999999 s
Default value
52 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
0.1 ... 999999.0 s
Default value
2.0 s
Stop measured value
Stops when the preset measured value has been reached since the start of
the measurement.
Range
0 ... 1E+99
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:
Measuring time
Maximum period of time for measurement.
Range
0 ... 999999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list. Max.
number of entries?
Range
0.1 ... 999999.0 s
Default value
2.0 s
Stop measured value
Stops when the preset measured value has been reached since the start of
the measurement.
Range
0 ... 1E+99
Selection
off
Default value
off
off
No stop.