5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

752

■■■■■■■■

tiamo 2.1

855
Selection

1 | 2 | 3

Default value

1

Titrino
Selection

internal D0

736, 751, 799
Selection

internal D0 | external D1 | external D2

Default value

internal D0

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando (without 852)
Selection

1 | 2

Default value

1

852
Selection

2

Default value

2

Sensor

Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode